Skip to product information
1 of 1

Labman

Automatic Density Meter ADM5CFR (0 to 3.00000 g/cm³)

Automatic Density Meter ADM5CFR (0 to 3.00000 g/cm³)

Regular price Rs. 532,500.00
Regular price Sale price Rs. 532,500.00
Sale Sold out
Shipping calculated at checkout.

Low stock: 10 left

The ADM5CFR is a high-end Automatic Density Meter designed for industries requiring the highest levels of precision and regulatory compliance. Operating on the principle of U-tube vibration with electronic excitation, this model offers an exceptional density resolution of 0.00001 g/cm³. It is specifically engineered for the pharmaceutical, petroleum, and chemical industries where data integrity is paramount.

The ADM5CFR is fully compliant with 21 CFR Part 11 regulations, providing robust audit trails, user management permissions, and approved electronic signature capabilities. It features an integrated Peltier temperature control system with a wide range (5°C to 70°C), a 10.1-inch Android touch interface, and a 1080P HD camera for real-time sample observation and bubble detection. With over 30 built-in measurement units and the ability to export professional PDF/Excel reports, it is the definitive solution for regulated laboratory environments.

Key Features 

Automatic Density Meter function is to quickly measure the density of liquids

Quartz glass U-Shaped Tube and photoelectric detection methods

High definition camera video for detection

Minimum sample 2~3 ml for each measurements

Built-in Peltier Temperature control

Compatible with Manual sampling and automatic sampling

10.1" Touch screen display with an Android operating system.

1080P high-definition camera for observing the sample injection process

Bubble Detection users observe and identify bubbles or impurities in the sample.

Measurement results can be saved as PDF reports and exported via USB

Batch data collection and storage

RS232C Serial port, USB

Air pump for drying and cleaning

Print results by thermal printers (Optional)

About 21CFR Part 11 - ADM5CFR: Defines requirements for submitting documents in electronic form and the criteria for approved electronic signatures.

Product Specification Table

Specification Detail
Model ADM5CFR
Density Measuring Range 0.00000 to 3.00000 g/cm³
Density Resolution 0.00001 g/cm³
Density Accuracy ±0.0001 g/cm³
Density Repeatability ±0.00005 g/cm³
Temperature Control Range 5°C ~ 70°C
Temperature Repeatability ± 0.02°C
Temperature Accuracy ±0.05°C
Communication Port RS232 Interface, USB, RJ45
Measurement Units g/cm³, kg/m³ (More than 30 built-in units)
Audit Trial, User management Yes (21 CFR Part 11)
Temperature Control Peltier (In-Built)
Display 10.1" Touch screen
Observation 1080P HD camera
Data Storage 128GB
Measurement Records 2,00,000 groups
Operating System Android
Report Pdf / Excel
Power Supply AC 220V, 50/60Hz
Dimension (LxWxH) 730 x 480 x 420 mm
Gross Weight 27kg
View full details